Ultracentrifuges Optima XL Ultracentrifuge Tabletop ultracentrifuge, Optima TL Contact: Sonja Young Rm: NSC 518 Tel: (404) 413-5379 |
![]() |
|
Attention!!! If ultracentrifuge not working for any reason, press clear and try again. General Warnings: |
| Optima XL Ultracentrifuge 1. All tubes must be as closely balanced as you can get them. 2. Make sure all quick-seal tubes are properly sealed. To check seal, hold sealed end pointing away from your face, gently squeeze the tube. There shouldn’t be any oozing. If there is, try to reseal. If this doesn’t work (and it usually doesn’t), put sample in a new tube and try again. Failure to check for a proper seal, usually ends with the loss of the sample. Improperly sealed tubes collapse during a run and get all over the rotor. 3. Log in run in log book. 4. The power should be on. Standby mode the screen is blank. Selecting any key will return the screen to Operating mode. The centrifuge has a circuit breaker instead of a power switch. I = on and O = off; located on the right side of the panel. 5. The screen has the current values on the left and the settings on the right. It also has some status information along the bottom (for example, whether the vacuum is on or off. 6. Make sure the overspeed disk is on the bottom of your rotor. This prevents the rotor from going too fast. 7. Open door and place rotor in chamber. Make sure it is seated properly. 8. Press speed. The Speed field blinks. Use the keypad to change the speed. Press enter. 9. Press time. The Time field blinks. Use the keypad to change the time. Press enter. 10. Press temp. The Temp field blinks. Use the keypad to change the time. Press enter. 11. 2 of the softkeys are for acceleration and deceleration. Usually you want maximum acceleration, which is 9. For different applications, you will want different deceleration speeds. You can select No Brake or 1 – 9, with 1 the slowest deceleration. 12. Start run. The vacuum will automatically start. The rotor will stop at 3000 rpm until the vacuum reaches 750 microns and then will continue on to the set speed. 13. Come back in 15 or 20 min. to make sure the run is going as you expect – temperature correct, still holds vacuum and the speed is what you set. 14. At the end of the run, the vacuum must be vented. Press vacuum to release the vacuum. Open door to remove rotor. 15. If there are ANY spills in the centrifuge, clean the drum out with water. If there are ANY spills or collapsed tubes, clean out the rotor and air dry completely. Beckman Tabletop ultracentrifuge, Optima TL 1. Power on. Log in run in log book. This information is given as a guide to the facilities and instrumentation available in the DNA/Protein Core facility at Georgia State University. If you have any concerns or thoughts about the content of this website please contact: John Houghton (404) 413-5390
|